среда, июля 09, 2008

826-Mhz and 125-kHz RFID induce EMI incidents

Наконец была озвучена проблема электромагнитной совместимости RFID-ридеров с аппаратурой применяемой в больницах .
Приговор завис над 900 Мгц и 125 кГц ридерами которые своим излучением могут вызвать сбои например в кардиостимуляторах. Выдержка из отчета Эрика Ян ван Лишаута (Erik Jan van Lieshout) :

Results In 123 EMI tests (3 per medical device), RFID induced 34 EMI incidents: 22 were classified as hazardous, 2 as significant, and 10 as light. The passive 868-MHz RFID signal induced a higher number of incidents (26 incidents in 41 EMI tests; 63%) compared with the active 125-kHz RFID signal (8 incidents in 41 EMI tests; 20%); difference 44% (95% confidence interval, 27%-53%; P < .001). The passive 868-MHz RFID signal induced EMI in 26 medical devices, including 8 that were also affected by the active 125-kHz RFID signal (26 in 41 devices; 63%). The median distance between the RFID reader and the medical device in all EMI incidents was 30 cm (range, 0.1-600 cm).
Понятно почему не тестировались ридеры работающие на 13.56 Мгц. В маломощном варианте они не замечаются на расстояниях более 10-15 см даже самими RFID-метками.

Источники:

Radio tag signals can play havoc with hospital devices - health - 24 June 2008 - New Scientist

http://www.timedomain.com/about/Newsroomcontent/JAMAPublishesReportonRFID.pdf

Комментариев нет: